Cracks may occur over time on the tube holding shaft near the base of the tube mounting flange.
Source facts
Action
Shimadzu sent an Urgent Voluntary Medical Device Recall Notice to all affected customers. The firm will implement a voluntary recall/field corrective action to add protective parts preventing the x-ray tube assembly from coming into contact with the patient or operator should separation of the tube mounting flange from the tube holding shaft occur. Customers with questions were instructed to call 800-228-1429.
Initiated
2017-05-23
Root cause
Device Design
Terminated
2019-10-01
Distribution
Worldwide Distribution - Nationwide distribution to AK, AR, AZ, CA, CT, FL, IL, IN, KS, LA, MI, MO, NE, NJ, NM, NV, NY,OH, OK, PA, RI, SC, TN, TX, WA, WI, and Hawaii
Foreign: Canada and Japan
Recall number
Z-2653-2017
Recall status
Terminated
CDRH record ID
155806
FDA record created
2017-06-29
Recall event number
77351