Skip to content

Public FDA record

Z-2653-2023: Apellis Injection Kit 29g Injection Needle -single use intended to inject fluids into, or withdraw fluids

Recall record Z-2653-2023 for Apellis Pharmaceuticals, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2653-2023
Company in FDA record
Apellis Pharmaceuticals, Inc.
FDA classification date
Sep 29, 2023
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

19-gauge x 1" inch filter needles showed appearances of a small fraction of non-preferred characteristics. A causal relationship has not been established between the structural variations in this 19-gauge filter needle and the rare events of retinal vasculitis

Source facts

Action
Apellis issued Medical Device Field Correction letters via email on 8/22/23 to Healthcare Practitioners and Distributors/Specialty Distributors and Pharmacy/ Logistics. The communication to the Healthcare Practitioners also includes update on the rare events of retinal vasculitis with real world use of SYFOVRE (pegcetacoplan injection). A causal relationship has not been established between the 19-gauge needle and the rare events of retinal vasculitis in the real world. Letter states reason for recall, health risk and action to take: Actions to be taken by the Customer/Distributor: Please note that as part of this field action, you are requested to: " Confirm your availability and participation by sending back an email to acknowledge receipt of this message. " Immediately cease distribution of the below injection kit lots and place any remaining inventory of the lots On Hold. " Provide a summary of current on-hand inventory of the injection kit lots below. " Provide a distribution report of the injection kit lots including customer name/address, lot shipped, quantity shipped, and date. Please provide the Distribution report in a maximum time of four business hours. Apellis will also be recommending that practitioners discontinue use of, and dispose, any remaining injection kits that contain the specific 19-gauge filter needle and recommend they instead use the injection kits containing the 18-gauge filter needle already in distribution. Apellis is now exclusively distributing injection kits with the 18-gauge filter needle. If you need to place a supplemental order to replace quarantined lots, please place them immediately through your normal ordering process and Apellis will expedite shipping replacement injection kits./ Apellis will also be notifying physicians that if they do not have an injection kit with the 18-gauge filter needle, they may request a new kit at no cost by emailing info@apellis.com or contacting their Apellis Territory Business Mana
Initiated
2023-08-22
Root cause
Under Investigation by firm
Distribution
US Nationwide distribution.
Recall number
Z-2653-2023
Recall status
Open, Classified
CDRH record ID
202936
FDA classified
2023-09-29
Recall event number
92926

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗