Skip to content

Public FDA record

Z-2653-2026: Covidien Shiley Disposable Decannulation Plug REF DDCP UDI-DI code: 10884522004060 Product Codes that in

Recall record Z-2653-2026 for Covidien LLC. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2653-2026
Company in FDA record
Covidien LLC
FDA classification date
Jul 2, 2026
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Due to product labeling incorrectly including the symbol indicating non-DEHP (Bis(2-ethylhexyl) phthalate as the product includes 30 to 35% DEHP.

Source facts

Action
On 04/23/2026, an "Urgent Medical Device Correction" letter was mail/email to customers informing them that the labeling for all lots of Covidien Shiley" disposable decannulation plugs incorrectly includes the symbol indicating non-DEHP (Bis(2-ethylhexyl) phthalate). The seal of this product includes 30 to 35% DEHP. Customers are instructed to: -No additional action is required by clinicians for the care or follow-up of patients who are using or have used Covidien Shiley" disposable decannulation plugs. -Customers can continue to use the Covidien Shiley" disposable decannulation plugs. -Post this notice in all care environments where the affected products are used to maintain awareness to this labeling issue. -Share this notice with all relevant personnel within their organization, and with any organization to which the affected product has been transferred or distributed. -Complete and return the enclosed Customer Confirmation Form to rs.gmbmitgfca@medtronic.comeven if you do not have unused inventory. Any questions regarding this communication, contact Medtronic Representative or Customer Service at 800-962-9888.
Initiated
2026-04-23
Root cause
Labeling Change Control
Distribution
U.S. Nationwide distribution including in the states and territories of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, RI SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, and Virgin Islands. The countries of Albania, Andorra, Argentina, Australia, Austria, Belgium, Bhutan, Canada, Canary Islands, Chile, Colombia, Costa Rica, Croatia, Cyprus, Czech Republic, Denmark, Dominican Republic, Finland, France, Georgia, Germany, Greece, Guadeloupe, Guyana, Hungary, Ireland, Israel, Italy, Japan, Kazakhstan, Kuwait, Latvia, Lithuania, Luxembourg, Martinique, Mexico, Netherlands, New Caledonia, New Zealand, North Macedonia, Norway, Oman, Paraguay, Peru, Poland, Portugal, Qatar, Reunion, Romania, Saudi Arabia, Slovakia, South Africa, Spain, Sweden, Switzerland, Togo, Trinidad And Tobago, Turkey, Uganda, United Arab Emirates, Uruguay.
Recall number
Z-2653-2026
Recall status
Open, Classified
CDRH record ID
219885
FDA classified
2026-07-02
Recall event number
98819

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.