Z-2654-2025: Biograph mMR. Model Number: 10433372.

Recall record Z-2654-2025 for Siemens Medical Solutions USA, Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2654-2025
Company in FDA record
Siemens Medical Solutions USA, Inc
FDA classification date
Sep 30, 2025
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class I

Reason stated in the source

There is a potential for an ice blockage to form or currently exist within the magnet venting system. In the event of a quench, helium gas may be unable to escape through the designed venting paths, leading to a pressure build-up within the helium containment system. This pressure build-up could ultimately rupture the helium containment system, potentially resulting in a helium leak into the scanning room.

Source facts

Action
Siemens Healthcare notified consignees on about 08/28/2025 via letter. Consignees were informed that Siemens Customer Service Organization will conduct an inspection of affected units to address the issue. If customers identify issues, they are cautioned to not reboot the system, to call their local Siemens Healthineers service organization, and grant access to the system. Siemens will inspect devices and perform a correction, if needed. Consignees were also instructed to ensure all users or customers are notified of the issue and to maintain the Customer Safety Advisory Notice. Consignees were also requested to confirm receipt and acknowledgement of the notification. Siemens expanded the recall and notified consignees on about 10/01/2025 via letter. These notifications included the same information as initially sent to consignees. A follow up communication was sent to consignees on about 10/15/2025 via letter. This informed users of further details and additional necessary safety actions. If a specific "Magnent Supervision" error message appears on the host computer, no further scans can be performed, and you are requested to: not reboot the system, remove the patient from the system, lock/secure all access to the scanning room with the system remaining energized at field, post a DO NOT USE notice on or near the device, and call the Siemens Healthineers service organization. In addition, Siemens reminded users that the Magnet Stop button is only intended for emergencies. DO NOT initiate a manual quench of the magnet unless there is an emergency. Ensure all designated emergency exit routes remain unobstructed and are marked. Consignees were also instructed to ensure all users or customers are notified of the issue, to maintain the Customer Safety Advisory Notice is retained in product related records, to confirm receipt and acknowledgement of the notification, and if the affected system is no longer in the consignee's possession, notify the new owner. ***Updat
Initiated
2025-08-28
Root cause
Under Investigation by firm
Distribution
Worldwide distribution - US Nationwide and the countries of United Arab Emirates, Armenia, Angola, Argentina, Austria, Australia, Azerbaijan, Bosnia and Herzegovina, Bangladesh, Belgium, Bulgaria, Bahrain, Brunei Darussalam, Brazil, Botswana, Belarus, Canada, Switzerland, Chile, China, Colombia, Cyprus, Czechia, Germany, Denmark, Algeria, Ecuador, Estonia, Egypt, Spain, Ethiopia, Finland, Faroe Islands, France, United Kingdom of Great Britian and Northern Ireland, Georgia, Greece, Hong Kong, Croatia, Hungary, Canary Islands, Indonesia, Ireland, Israel, India, Iran, Iceland, Italy, Jordan, Japan, Kenya, Korea, Kuwait, Kazakhstan, Lebanon, Sri Lanka, Lithuania, Luxembourg, Latvia, Morocco, Monaco, Moldova, Montenegro, Republic of North Macedonia, Myanmar, Mongolia, Macao, Malta, Maldives, Mexico, Malaysia, Mozambique, Nigeria, Netherlands, Norway, Nepal, New Zealand, Oman, Panama, Peru, Philippines, Pakistan, Poland, Portugal, Paraguay, Qatar, Reunion, Romania, Serbia, Russian Federation, Saudi Arabia, Sweden, Singapore, Slovenia, Slovakia, Senegal, Syrian Arab Republic, Thailand, Turkmenistan, Tunisia, Turkey, Trinidad and Tobago, Taiwan, Tanzania, Ukraine, Uganda, Uzbekistan, Viet Nam, Kosovo, South Africa.
Recall number
Z-2654-2025
Recall status
Open, Classified
CDRH record ID
215853
FDA classified
2025-09-30
Recall event number
97530

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗