Skip to content

Public FDA record

Z-2656-2023: F2-01 Frame, physiological patient monitor (with arrhythmia detection or alarms)

Recall record Z-2656-2023 for GE Medical Systems China Co., Ltd.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2656-2023
Company in FDA record
GE Medical Systems China Co., Ltd.
FDA classification date
Sep 29, 2023
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

There is a potential interruption of data communication between E-modules inserted in the F2-01 Frame and CARESCAPE ONE and CARESCAPE Canvas 1000 patient monitors if the F2-01 Frame has not been powered down within the last 120 days.

Source facts

Action
The consignee recall notification was sent out via FedEx on 09/08/2023. The letter instructs the consignee that they can continue to use the F2-01 Frame by completing the power down instructions located within the letter at least every 90 days, until GE HealthCare has corrected the affected device. The consignee is also requested to ensure all potential users in your facility are made aware of this safety notification and the recommended actions; retain the document for their records; and to complete and return the attached acknowledgement form to Recall.36160@ge.com.
Initiated
2023-09-08
Root cause
Software design
Distribution
Worldwide distribution - US Nationwide and the countries of Australia, Austria, Bahrain, Belgium, Bulgaria, Cyprus, Czech Republic, Finland, France, Germany, Ireland, Italy, Kuwait, Netherlands, New Zealand, Oman, Poland, Qatar, Saudi Arabia, Slovakia, Slovenia, Spain, Switzerland, Turkey, United Arab Emirates, and United Kingdom.
Recall number
Z-2656-2023
Recall status
Open, Classified
CDRH record ID
203364
FDA classified
2023-09-29
Recall event number
93054

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗