Z-2656-2026: CK000495A Poly Midline Catheter 4 Fr Single-Lumen UDI-DI Code: 00801741121609 CK000566 Poly Midline Cath

Recall record Z-2656-2026 for Bard Access Systems, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2656-2026
Company in FDA record
Bard Access Systems, Inc.
FDA classification date
Jul 2, 2026
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Due to lidocaine ampoules being manufactured under deficient manufacturing practices by manufacturer, lidocaine ampoules are currently being recalled by supplier.

Source facts

Action
On 05/21/2026, the firm FedEx'd an "URGENT: Medical Device Recall" letter informing customers/distributed that BD has become aware that the FDA has placed Lidocaine manufactured by Huons Co., Ltd. on import alert #66-40 - Detention Without Physical Examination of Drugs From Firms Which Have Not Met Drug GMPs. BD is recalling products that contain the affected lidocaine ampules. Customers are instructed to: 1. Not to use the affected Lidocaine ampules. 2. Secure an alternative local Lidocaine drug prior to entering the procedural environment. 3. Upon opening the kit, remove and destroy the drug product from the BD convenience kits/procedure trays at the point of use, within a controlled and sterile environment to maintain the sterility of the remaining components within the kit. All other kit/tray components are safe to use. 4. If affected Lidocaine ampule was previously used without issue, no additional treatment or clinical follow up is required. 5. If affected Lidocaine ampule was previously used with issue, report the incident per institutional guidelines and take any necessary actions. Additional actions to take: 1. Circulate this notice to all those who need to be aware within your organization or to any organization where the potentially affected products have been transferred. 2. Post this notice on all inventory storage locations and any location where the product may be used. 3. Label affected kits/trays using the template provided in Appendix 2 to facilitate easy identification of products containing the affected ampules. The label can be printed using Avery template 5160. a. Apply the labels to the product without covering or obscuring any regulatory or traceability information. b. Apply to the top web of the kit/tray pressing firmly to ensure adhesion, while taking care not to damage or compromise the sterile barrier. For Questions - Contact North American Regional Complaint Center at 1-844-8BD-LIFE (1-844-823-5433) Say Product Complaints when prompt
Initiated
2026-05-21
Root cause
Nonconforming Material/Component
Distribution
Worldwide U.S. Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, RI, SC, SD, TN, TX, UT, VA, WA, WI, WV, and WY. The country of Belgium.
Recall number
Z-2656-2026
Recall status
Open, Classified
CDRH record ID
220785
FDA classified
2026-07-02
Recall event number
99096

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗