Open original FDA record ↗Recall record
- FDA identifier
- Z-2657-2011
- Company in FDA record
- Arcadia Medical Corporation
- FDA classification date
- Jun 24, 2011
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 17, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
The product does not contain a punched hole above the cuff access port (a lumen that enables suctioning of secretions proximal to the inflated cuff) as intended by the devices design. This eliminates the user's ability to suction secretions through this suction port, necessitating that the user revert to standard of care methods for suctioning secretions from around the cuffed trach tube shaft.
Source facts
- Action
- Arcadia Medical Corporation sent a "MEDICAL DEVICE RECALL" letter on May 9, 2011 and email on May 8, 2011 to all affected costumers. The letter included; description of the product name, afffected lot numbers, description of the problem, instructions to discontinue use of the recalled devices and to return them to Arcadia Medical for replacement . The letter instructed the customers to email Arcadia Medical with the number of devices in their inventory. For additional information contact Arcadia Medical Corporation at 219-779-9431.
- Initiated
- 2011-05-08
- Root cause
- Process control
- Terminated
- 2011-08-30
- Distribution
- Worldwide Distribution --USA including the state of Ohio, and the countries of Denmark, Italy, Germany, Turkey, Slovakia, Isreal, France, Sweden, and Romania
- Recall number
- Z-2657-2011
- Recall status
- Terminated
- CDRH record ID
- 100382
- FDA classified
- 2011-06-24
- Recall event number
- 58828