Motion Concepts identified that there is a possibility of knee support failure occurring through fatigue of two attachment bolts in the Vertical Position System (TRx Vertical Positioner).
Source facts
Action
URGENT RECALL NOTIFICATION Letters (dated 8/26/2014) were sent to customers on 8/26/14 via Federal Express. Motion Concepts is requesting that all customers that have purchased a Motion Concepts TRx Vertical
Positioner system in between June 2011 - May 2014 to make arrangements for replacement of the knee support mount in accordance with the instructions enclosed in the letter. For questions, please contact us at 905-695-0134.
Initiated
2014-08-28
Root cause
Device Design
Terminated
2015-06-03
Distribution
US Distribution in states of: AR, CA, CT, DE, IA, ID, IL, IN, MI, MD, MO, NE, NY, OH, PA, TN, TX, and WA.
Recall number
Z-2657-2014
Recall status
Terminated
CDRH record ID
129660
FDA classified
2014-09-17
Recall event number
69113