Open original FDA record ↗Recall record
- FDA identifier
- Z-2657-2017
- Company in FDA record
- CardioTek BV
- FDA record creation date
- Jun 30, 2017
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Software bug which allows parameters to be changed unintentionally during use.
Source facts
- Action
- The Firm, Schwarzer cardiotek, sent an "Urgent Field Safety Notice" dated 4/21/2017 to its consignees. The notice described the product, problem and actions to be taken. The consignees were instructed to do the following: " While backup pacing is active the software must not be operated. " In case of backup pacing disconnect all catheter connection pins and reconnect only the two required for stimulation to the dedicated backup pacing output ports (OUT1 or OUT2). " If the backup pacing parameters were changed by accident quickly switch backup pacing off and on again to reset the parameters. " Ensure that the system is only operated by trained users. The firm plans to do the following: " provide an amendment to the hardware manual concerning the correct use of the backup pacing functionality to all service partners and customers. An updated software, which prevents accidental change of backup pacing parameters will be made available. Upon receipt of the amendment and installation of the software update, the completed feedback form must be returned to CardioTek no later than 3/4/2018 by fax +49 7131 2774 590 or email: Service@schwarzercardiotek.com. If you have any questions, please contact Sales & Service at +49 172 790 9112 or email: Service@schwarzercardiotek.com.
- Initiated
- 2017-04-13
- Root cause
- Software design
- Distribution
- Worldwide Distribution: US (nationwide) to states of: CA, IL, OH, and MN; and countries of: Albania, Algeria, Argentina, Bangladesh, Belgium, Bolivia, Brazil, Canada, Chile, Colombia, Costa Rica, Croatia, Cuba, Czech Rep, Denmark, Ecuador, Egypt, Estonia, Finland, France, Georgia, Germany, Greece, Guatemala, Honduras, Hungary, India, Indonesia, Iran, Iraq, Ireland, Israel, Italy, Jamaica, Jordan, Kosovo, Kuwait, Latvia, Lebanon, Libya, Lithuania, Malaysia, Malta, Mexico, Monaco, Morocco, Netherlands, Norway, Pakistan, Palestine, Peru, Philippines, Poland, Portugal, Rep. Dominicana, Romania, Russia, Saudi Arabia, Singapore, South Africa, Spain, Sri Lanka, Sweden, Switzerland, Syria, Trinidad & Tobago, Turkey, United Arab Emirates, United Kingdom, Venezuela and Vietnam.
- Recall number
- Z-2657-2017
- Recall status
- Completed
- CDRH record ID
- 154842
- FDA record created
- 2017-06-30
- Recall event number
- 77025