Z-2658-2023: Heartmate 3 LVAS Implant Kit, REF: 106524US, 106524INT; Heartmate XVE & HeartMate II Apical Coring Knife,

Recall record Z-2658-2023 for Thoratec Corp.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2658-2023
Company in FDA record
Thoratec Corp.
FDA classification date
Sep 29, 2023
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Inability to start and/or complete the coring procedure due to inadequate sharpness of the Apical Coring Knife, provided with Left Ventricular Assist System, could result in an extended procedure or surgical time while the backup coring knife is acquired.

Source facts

Action
On 8/7/23 customers were notified of the recall verbally, via email, in person visits, and phone calls. 8/21/23 the Abbott sales representatives started hand delivering correction notices to customers were told the firm was reinforcing the following warning in the Instructions for Use: 1) During the implant process, a complete backup system (implant kit and external components) must be available on-site and in close proximity for use in the event of an emergency. 2) Additionally, if unusual resistance is encountered during the coring procedure, immediately stop coring and complete using the backup Apical Coring Knife. Customers were also informed that in cases were the recall issue occurred during use, the physician was able to complete the procedure by using an Apical Coring Knife from the backup LVAS implant kit or an alternate surgical tool. Distribute this notice to those who need to be aware within your institution. Firm is continuing its investigation on this matter and working diligently to resume normal supply. Direct questions to your local Abbott representative. Complete and return the acknowledgement form via email to MCSHMNotices@Abbott.com Correction notices are continuing to be distributed identifying additional affected serial numbers. The firm is continuing to work with the apical coring knife vendor to supply knives that do not present this issue. However, the supply of unaffected apical coring knives continues to be limited.
Initiated
2023-08-07
Root cause
Process control
Distribution
Worldwide - US Nationwide distribution including in the states of AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, IA, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, NC, NE, NJ, NM, NV, NY, OH, OK, OR, PA, SC, TN, TX, UT, VA, WA, WI, WV and the countries of Australia, Austria, Belgium, Brazil, Bulgaria, Canada, Chile, Colombia , Czech Republic, Denmark, Egypt, Estonia, Finland, France, Germany, Greece, Hong Kong, India, Iran, Iraq, Ireland, Israel, Italy, Japan, Korea, Kuwait, Lebanon, Lithuania, Macedonia, Malaysia, Mexico, Morocco, Netherlands, New Zealand, Norway, Poland, Portugal, Qatar, Romania, Russia, Saudi Arabia, Serbia, Singapore, Slovenia, South Africa, Spain, Sweden, Switzerland, Taiwan, Thailand, Tunisia, Turkey, United Arab Emirates, United Kingdom.
Recall number
Z-2658-2023
Recall status
Open, Classified
CDRH record ID
203000
FDA classified
2023-09-29
Recall event number
92951

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗