Open original FDA record ↗Recall record
- FDA identifier
- Z-2659-2011
- Company in FDA record
- Pro Trainers' Choice Company
- FDA classification date
- Jun 24, 2011
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 17, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
The control panel of the Precision Ice One Patient Use Control Unit and Wrap System has a overheating issue. The control panel printed circuit boards were missing a jumper connection across resistors R28 A, B, C, and D.
Source facts
- Action
- Pro Trainers' Choice sent a Mandate-Medical Device Recall on April 18, 2011 and April 20, 2011 to all affected costumers. The letter included a description of the product, affected model and lots. It instructed the costumers to discontinue using the Precision Ice, Model 4BBK02BK until the device is returned and corrected by Pro Trainers' Choice. The device should be returned to PO Box 27, Kingston, WA 98346. The customers were instructed to notify their sub-account customers of this recall if the devices were further distributed. Customers can call Pro Trainers' Choice Company at 360-297-3902 or visit the website www.precisionice.com for any questions about this recall.
- Initiated
- 2011-04-18
- Root cause
- Other
- Terminated
- 2012-03-01
- Distribution
- Nationwide distribution including CA, CO, FL, IN, KS, NY, OR, TX, and WI.
- Recall number
- Z-2659-2011
- Recall status
- Terminated
- CDRH record ID
- 100276
- FDA classified
- 2011-06-24
- Recall event number
- 58770