Skip to content

Public FDA record

Z-2659-2011: Precision Ice One-Patient-Use Control Unit and Wrap System (Model # 4BB02BK). The device is intended to

Recall record Z-2659-2011 for Pro Trainers' Choice Company. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2659-2011
Company in FDA record
Pro Trainers' Choice Company
FDA classification date
Jun 24, 2011
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 17, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

The control panel of the Precision Ice One Patient Use Control Unit and Wrap System has a overheating issue. The control panel printed circuit boards were missing a jumper connection across resistors R28 A, B, C, and D.

Source facts

Action
Pro Trainers' Choice sent a Mandate-Medical Device Recall on April 18, 2011 and April 20, 2011 to all affected costumers. The letter included a description of the product, affected model and lots. It instructed the costumers to discontinue using the Precision Ice, Model 4BBK02BK until the device is returned and corrected by Pro Trainers' Choice. The device should be returned to PO Box 27, Kingston, WA 98346. The customers were instructed to notify their sub-account customers of this recall if the devices were further distributed. Customers can call Pro Trainers' Choice Company at 360-297-3902 or visit the website www.precisionice.com for any questions about this recall.
Initiated
2011-04-18
Root cause
Other
Terminated
2012-03-01
Distribution
Nationwide distribution including CA, CO, FL, IN, KS, NY, OR, TX, and WI.
Recall number
Z-2659-2011
Recall status
Terminated
CDRH record ID
100276
FDA classified
2011-06-24
Recall event number
58770

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗