Skip to content

Public FDA record

Z-2662-2017: Logic Fit Tibial Tamp Head

Recall record Z-2662-2017 for Exactech, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2662-2017
Company in FDA record
Exactech, Inc.
FDA classification date
Jun 19, 2017
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

The firm determined that the labeling (surgical technique) should be updated to clarify the proper technique and instrumentation to remove the tamp assembly from the bone, i.e., to include a caution statement about the potential for instrument breakage if the tamp handle/guide is misused by impacting in retro-grade.

Source facts

Action
Consignees were sent a Field Advisory Notice dated 5/22/2017 by mail informing them updated surgical techniques are available on Exactech's website (www.exac.com). The notice requested consignees: Extend the information to accounts in possession of the product Complete & return the attached Field Advisory Notice Response Form to Exactech
Initiated
2017-05-25
Root cause
Error in labeling
Terminated
2019-07-26
Distribution
AL, AZ, CA, CA, CO, FL, GA, IA, ID, IL, IN, KS, LA, MA, MD, ME, MN, NC, NJ, NV, NY, OH, OK, OR, RI, SC, TN, TX, VA, WA, WI, Hawaii and Puerto Rico Australia, France, Germany, India, Italy, Japan, Korea, Luxembourg, Singapore, Spain, Switzerland, The Netherlands and United Kingdom
Recall number
Z-2662-2017
Recall status
Terminated
CDRH record ID
154803
FDA classified
2017-06-19
Recall event number
77001

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗