Open original FDA record ↗Recall record
- FDA identifier
- Z-2662-2017
- Company in FDA record
- Exactech, Inc.
- FDA classification date
- Jun 19, 2017
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
The firm determined that the labeling (surgical technique) should be updated to clarify the proper technique and instrumentation to remove the tamp assembly from the bone, i.e., to include a caution statement about the potential for instrument breakage if the tamp handle/guide is misused by impacting in retro-grade.
Source facts
- Action
- Consignees were sent a Field Advisory Notice dated 5/22/2017 by mail informing them updated surgical techniques are available on Exactech's website (www.exac.com). The notice requested consignees: Extend the information to accounts in possession of the product Complete & return the attached Field Advisory Notice Response Form to Exactech
- Initiated
- 2017-05-25
- Root cause
- Error in labeling
- Terminated
- 2019-07-26
- Distribution
- AL, AZ, CA, CA, CO, FL, GA, IA, ID, IL, IN, KS, LA, MA, MD, ME, MN, NC, NJ, NV, NY, OH, OK, OR, RI, SC, TN, TX, VA, WA, WI, Hawaii and Puerto Rico Australia, France, Germany, India, Italy, Japan, Korea, Luxembourg, Singapore, Spain, Switzerland, The Netherlands and United Kingdom
- Recall number
- Z-2662-2017
- Recall status
- Terminated
- CDRH record ID
- 154803
- FDA classified
- 2017-06-19
- Recall event number
- 77001