Open original FDA record ↗Recall record
- FDA identifier
- Z-2662-2018
- Company in FDA record
- Smiths Medical ASD Inc.
- FDA record creation date
- Aug 8, 2018
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Certain Barrel Clamp Guides may contain a ridge that could potentially lead to spring slippage, resulting in the inability of the pump to recognize a loaded syringe or the pump, misidentifying the size of a loaded syringe.
Source facts
- Action
- Customers were notified of the recall on about 04/13/2018 via URGENT MEDICAL DEVICE RECALL NOTICE letter. Instructions included to evaluate current inventory for affected Barrel Clamp Guide, review and complete the Recall Response Form, return affected Barrel Clamp Guides to Stericycle for destruction, and if the affected Barrel Clamp Guides have already been installed in pumps, continue to use the pumps, but utilize the Syringe Verification Reference Tool provided with the Recall Notice until the pumps are repaired. Customers were also instructed to notify customers if the affected devices were further distributed. Questions regarding the recall can be directed to Stericycle at SmithsMedical5120@stericycle.com.
- Initiated
- 2018-04-13
- Root cause
- Process change control
- Terminated
- 2023-01-30
- Distribution
- US distribution to AL, AZ, CA, DC, FL, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, NJ, NM, NY, OH, OK, PA, TN, TX, UT, VA, WA. International distribution to Canada and Lebanon.
- Recall number
- Z-2662-2018
- Recall status
- Terminated
- CDRH record ID
- 165595
- FDA record created
- 2018-08-08
- Recall event number
- 80490