Open original FDA record ↗Recall record
- FDA identifier
- Z-2663-2016
- Company in FDA record
- Tecan US, Inc.
- FDA classification date
- Aug 25, 2016
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
The batteries have the potential to overheating, melt or char.
Source facts
- Action
- Tecan contacted all customers by Safety Announcement, via certified mail starting July 25, 2016. The letter identified the product, the problem, and the action to be taken by the customer. Because affected batteries pose a fire and burn hazard, it is extremely important that customers check whether their battery is affected. Customers were advised to go immediately to the HP Battery Recall website at: http://www.HP.com/go/batteryprogram2016 to check if their battery is affected by this recall. Customers were also advised to complete and return the Customer Response Form at QA@tecan.com or fax it to 919-287-2961. Customers with questions should contact HP at 1-888-202-4320 or Tecan Technical Service's Helpdesk at 800-832-2687, option 1.
- Initiated
- 2016-07-25
- Root cause
- Nonconforming Material/Component
- Terminated
- 2017-01-23
- Distribution
- Worldwide Distribution - US including AK, AL, AR, CA, FL, GA, HI, KY, MA, MD, NC, NY, PA, PR, SC, TN, TX, UT, VA. International account: Canada.
- Recall number
- Z-2663-2016
- Recall status
- Terminated
- CDRH record ID
- 148496
- FDA classified
- 2016-08-25
- Recall event number
- 74806