Z-2665-2023: therascreen KRAS RGQ PCR Kit (24)-IVD qualitative PCR assay used on the Rotor-Gene Q MDx instrument for t

Recall record Z-2665-2023 for Qiagen GmbH. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2665-2023
Company in FDA record
Qiagen GmbH
FDA classification date
Sep 29, 2023
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Handbook Revision (Rev13) a dilution step is indicated as an Action when having the Flag SAMPLE_INT_CTRL_FAIL. A dilution of CRC tissue could lead to a false negative result. To remove ambiguity, an alteration must be made to the S the sample result with this flag should be interpreted as invalid

Source facts

Action
Qiagen US issued Urgent Medical Device Correction Letter to US consignees -therascreen KRAS RGQ PCR Kit (24), REF 870021 on 8/22/23 via email. Letter states reason for recall, health risk and action to take: This notice only applies to you if you downloaded and used the current version (Revision 13, January 2023) of the Instructions for Use. If you downloaded and used the current version (Revision 13, January 2023) of the Instructions for Use as mentioned above, follow these actions: o The dilution step of CRC tissue sample is incorrectly indicated and should not be followed. Follow the correct Action as described in this notice. o For NSCLC tissue sample follow the corrected Action as described in this notice. " Forward this information to all individuals and departments within your organization who are using the therascreen KRAS RGQ PCR Kit (24), REF. 870021. " If you are not the end user, please forward this notice to the product end user. " Review this notice with your laboratory/medical director. " Complete the Acknowledgement of Receipt Form attached to this letter by September 01, 2023, and email it to quality.communications@qiagen.com. Actions taken by QIAGEN QIAGEN is working on updating the therascreen KRAS RGQ PCR Kit Instructions for Use . If you have any questions or concerns, please contact your local QIAGEN Technical Service Department through any of the following: TechService-NA@qiagen.com
Initiated
2023-08-22
Root cause
Under Investigation by firm
Distribution
Worldwide - US Nationwide distribution in the states of AZ, CA, FL, IL, IN, KS, MA, NC,NY,OR, PA, TN, TX and the countries of KR, CN
Recall number
Z-2665-2023
Recall status
Open, Classified
CDRH record ID
203179
FDA classified
2023-09-29
Recall event number
92991

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗