Skip to content

Public FDA record

Z-2670-2017: Jelco, Hypodermic needle Pro needles 19G X1 1/2" 800/CA Model No. EL 1915 & 19G X 2" 600/CA Model No. SB

Recall record Z-2670-2017 for Smiths Medical ASD Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2670-2017
Company in FDA record
Smiths Medical ASD Inc.
FDA record creation date
Jul 3, 2017
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Smiths Medical became aware on January 11, 2017, that syringe assemblies provided by Nipro Medical Corporation contained pin holes in the side wall of the hub. The defect was not detected during Smiths Medical incoming inspection. The syringe assemblies were used to manufacture seven (7) finished good lots of Hypodermic Needle Pro needles, Smiths Medical part numbers EL 1915 and SB5021.

Source facts

Action
Smiths Medical sent an "Urgent Medical Device Recall Notice" dated April 04, 2017. The letter was addressed to Quality Assurance Manager & Associate Sourcing Consultant. The letter listed the Affected devices described the Reason for Recall, & Risk to Health. Requested consignee to complete Urgent Medical Device Recall Response Form within 10 days of receipt and send it to: fieldactions@smithsmedical.com. For further questions, please call (603) 352-3812
Initiated
2017-04-04
Root cause
Process control
Terminated
2019-08-05
Distribution
US Distribution to the state of : IN
Recall number
Z-2670-2017
Recall status
Terminated
CDRH record ID
155955
FDA record created
2017-07-03
Recall event number
77402

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗