Open original FDA record ↗Recall record
- FDA identifier
- Z-2670-2017
- Company in FDA record
- Smiths Medical ASD Inc.
- FDA record creation date
- Jul 3, 2017
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Smiths Medical became aware on January 11, 2017, that syringe assemblies provided by Nipro Medical Corporation contained pin holes in the side wall of the hub. The defect was not detected during Smiths Medical incoming inspection. The syringe assemblies were used to manufacture seven (7) finished good lots of Hypodermic Needle Pro needles, Smiths Medical part numbers EL 1915 and SB5021.
Source facts
- Action
- Smiths Medical sent an "Urgent Medical Device Recall Notice" dated April 04, 2017. The letter was addressed to Quality Assurance Manager & Associate Sourcing Consultant. The letter listed the Affected devices described the Reason for Recall, & Risk to Health. Requested consignee to complete Urgent Medical Device Recall Response Form within 10 days of receipt and send it to: fieldactions@smithsmedical.com. For further questions, please call (603) 352-3812
- Initiated
- 2017-04-04
- Root cause
- Process control
- Terminated
- 2019-08-05
- Distribution
- US Distribution to the state of : IN
- Recall number
- Z-2670-2017
- Recall status
- Terminated
- CDRH record ID
- 155955
- FDA record created
- 2017-07-03
- Recall event number
- 77402