Skip to content

Public FDA record

Z-2671-2011: B. Braun Celsite Venous Access Ports , Distributed by Aesculpa Inc. Model ST301, ST305V, ST301G, ST501G

Recall record Z-2671-2011 for B. Braun Interventional Systems. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2671-2011
Company in FDA record
B. Braun Interventional Systems
FDA classification date
Jun 27, 2011
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 17, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

The affected products contain the incorrect winged Surecan needle accessory component in certain Celsite Access Port package configurations. The winged Surecan needle is used to deliver medications and fluids into implanted Celsite Access Ports. Two winged Surecan needle sizes are provided in Celsite U.S. package configurations: 20G (PN 4058631) or 22G (PN 16006593A). A 20G or 22G needle

Source facts

Action
B.Braun Interventional Systems Inc, (BIS) sent a "Urgent Medical Device Recall" letter on April 16, 2011 to all affected customers. The letter described the problem and the product involved in the recall. Informed consignees that their local B.Braun Interventional System Inc Product Specialists will contact them with the specifics for return/replacement of the product upon return of qualified removed product. They also requested that a Celsite Access Port Patient Implant Form be filled out for each patient implanted with one of the Celsite Access Ports from the affected lots. This Form to be provided to them by their local BIS Product Specialist. For additional information contact 800 523-9676 x4459.
Initiated
2011-04-13
Root cause
Process design
Terminated
2012-07-06
Distribution
NY, ND, OK ,PA
Recall number
Z-2671-2011
Recall status
Terminated
CDRH record ID
100140
FDA classified
2011-06-27
Recall event number
58718

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗