Z-2671-2016: Centricity Laboratory Core Lab System 4.1 The Centricity Laboratory System is intended to be an informat

Recall record Z-2671-2016 for GE Healthcare It. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2671-2016
Company in FDA record
GE Healthcare It
FDA record creation date
Aug 26, 2016
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

The Centricity Laboratory removes free-text notes on a master panel when the user deletes the slave panel. The deletion of a slave panel will (auto slash) the corresponding test (item) results on the master panel. Additionally, the displayed and HL7 results show neither the appropriate abnormal flag value nor the reference range values, when using large_num.g >7 to report numeric results.

Source facts

Action
GE Healthcare, sent an "URGENT MEDICAL DEVICE CORRECTION" letter dated Nov 20, 2012 to all affected customers. The letter described the product, problems and actions to be taken. The customers were instructed to follow the safety instructions and to ensure that all potential users in their facility are made aware of this safety notification and the recommended actions. The software has been modified to correct these defects. The modified software will be released and installed in user sites through GE Healthcare FMI #86071. Customers were instructed to contact their GE Service Representative with any questions regarding this issue. For other questions, contact GE Helpdesk: North America; (905) 479-1249 or 1-888-778-3375; email: tgg.nahelpdesk@med.ge.com Europe; +44 1603 877595; email: tgg.ukhelpdesk@med.ge.com Asia Pacific; +61 3 5327 4000; email: tgg.ozhelpdesk@med.ge.com. For questions regarding this recall call 847-277-5000.
Initiated
2012-11-20
Root cause
Software design
Terminated
2017-12-29
Distribution
Worldwide Distribution - US Distribution to GA, OH, NY, TX, TN ,MN and Internationally to India, Australia, Canada, England, Qatar, and Scotland.
Recall number
Z-2671-2016
Recall status
Terminated
CDRH record ID
117092
FDA record created
2016-08-26
Recall event number
64796

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗