Z-2671-2025: VITEK 2 Gram-negative AST Test Kits (Cards) containing Colistin formulation cs02n. VITEK 2 is an automat

Recall record Z-2671-2025 for bioMerieux, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2671-2025
Company in FDA record
bioMerieux, Inc.
FDA classification date
Sep 29, 2025
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Potential for false resistant results for Colistin (cs02n) with multi-drug-resistant Pseudomonas aeruginosa and Acinetobacter baumannii complex used in specific cards

Source facts

Action
On 06/27/2025, the firm sent an "URGENT FIELD SAFETY NOTICE" letter to customer informing them that VITEK 2 users, who are using VITEK 2 GN-AST cards with Colistin formulation, cs02n, have reported obtaining false resistant Colistin results with Acinetobacter baumannii complex and Pseudomonas aeruginosa isolates. Customers are required to: bioMrieux recommends you confirm any resistant colistin (cs02n) results for multidrug resistant Pseudomonas aeruginosa and Acinetobacter baumannii complex isolates via an alternate method for all existing and future AST cards with cs02n. If appropriate for your laboratory, create a custom bioART rule to show when resistant colistin (cs02n) results are obtained for Pseudomonas aeruginosa and Acinetobacter baumannii complex isolates (refer to the VITEK 2 Software User Manual for instructions). Based on individual facility policy, you may wish to perform retrospective review of resistant results. Distribute this information to all appropriate personnel in your laboratory, retain a copy in your files, and forward this information to all parties that may use this product, including others to whom you may have transferred the product. Complete the Acknowledgement Form in Attachment A and return it to your local bioMrieux representative to confirm receipt of this notice. It is important that you return the acknowledgement form to bioMrieux even if you determine that this product correction notice does not impact your facility. For questions or further assistance, contact the local bioMerieux Customer Service Representative.
Initiated
2025-06-26
Root cause
Under Investigation by firm
Distribution
International distribution in the countries of Albania, Armenia, Australia, Austria, Belgium, Bosnia, Bulgaria, China, Croatia, Cyprus, Czech Republic/Slovakia, France, Germany, Greece, Iberia, India, Israel, Italy, Japan, Lithuania, Macedonia, Malaysia/Brunei, Moldova, Netherlands, Pakistan, Poland, Portugal, Romania, Russia, Serbia, South Korea, Spain, Switzerland, Taiwan, Thailand, and Turkey.
Recall number
Z-2671-2025
Recall status
Open, Classified
CDRH record ID
215020
FDA classified
2025-09-29
Recall event number
97269

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗