Skip to content

Public FDA record

Z-2673-2016: Diamond Coated Tips and Burs are accessories to the air powered Bone drill (VARIOSURG with and without li

Recall record Z-2673-2016 for Nakanishi Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2673-2016
Company in FDA record
Nakanishi Inc.
FDA record creation date
Aug 26, 2016
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Through an error, the labeling failed to include the symbol or other text indicating that the diamond coated tips were single use only. A contributing factor to this error is that in other markets such as Japan, the diamond coated tips are multiple use and are not limited to single use only. The tips are stainless steel with a coating of fine diamond powder. The tips are sold non-sterile and are to be sterilized prior to first use. The tips were cleared as single use only. This problem was discovered during an ongoing, proactive, quality system improvement plan that was initiated by Nakanishi in October 2014. Because the correct instructions for sterilization are not available, it is possible that the devices might not be sterile prior to use. The absence of the single use statement introduces a risk of cross-contamination if the tips are reprocessed and reused.

Source facts

Action
Naknishi sent out an URGENT : MEDICAL DEVICE CORRECTION letter on July 18, 2016 describing the potential safety issue regarding their Diamond Coated Products. The firm is requesting customers take the following actions: 1. Examine their inventory immediately to determine if they have any product, both used and unused, bearing any of the Nakanishi Product Codes listed in List of Products enclosed with this letter. 2. Immediately dispose of any USED product according to applicable regulations. 3. Segregate any UNUSED product for return to Nakanishi. 4. In order for the firm to complete their records, they are requesting that customers complete the enclosed Correction Activity Sheet and FAX or Mail to: NSK America, Corp. 1800 Global Parkway, Hoffman Estates, IL 60192 US Facsimile: 1-800-838-9328 Customers with questions were advised to Telephone: 1-888-675-1675. NSKrecall@NSKDental.com For questions regarding this recall call 847-843-7664.
Initiated
2016-07-08
Root cause
Error in labeling
Terminated
2018-06-25
Distribution
Nationwide Distribution to Illinois, Florida and Georgia.
Recall number
Z-2673-2016
Recall status
Terminated
CDRH record ID
148154
FDA record created
2016-08-26
Recall event number
74666

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗