Open original FDA record ↗Recall record
- FDA identifier
- Z-2673-2020
- Company in FDA record
- Medtronic Xomed, Inc.
- FDA record creation date
- Jul 21, 2020
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
During internal testing execution of the next generation of Integrated Power Console (IPC) prototype it was noted that the M5 Microdebrider was rotating at a higher speed than the set value in the cosole.
Source facts
- Action
- On June 12th 2020 the firm send a notification to its consignees giving then the following instructions: Actions Required: 1. Immediately locate, quarantine, and return any software cards for Software 1898072 IPC Upgrade v2.7.3.0. Instructions for return are found on the Consignee Confirmation form. 2. Check the software version currently installed on your IPC Console and complete the Customer Confirmation Form. If your IPC has software v2.7.3.0 installed, do not use this console with M5 handpiece. Please record your console serial number and the software version currently loaded. 3. Return the completed Customer Confirmation Form via email to rs.entfca@medtronic.com or regular mail to: Medtronic-Xomed 6743 Southpoint Drive N Jacksonville, FL 32216 ATTN: Mike Tomberlin FCA Coordinator 4. Contact your Medtronic Field Representative to install replacement software Version 2.7.4.0 (CFN 1898073).
- Initiated
- 2020-06-12
- Root cause
- Software design
- Terminated
- 2023-04-20
- Distribution
- Worldwide distribution - US Nationwide distribution including in the states of AL, AK, AZ, AR, CA,CO, CT, FL, GA, HI, ID,IL, IN, IA, KS, KY, LA, MD, MA, MI, MN, MS, MO, NE, NV, NH, NJ, NY, NC, OH, OK, PA, RI, SC, TN, TX, UT, VA, WV, WA, WI. The countries of Australia, Austria, Belgium, Canada, China, Denmark, Finland, France, Germany, Greece, Italy, Japan, Luxembourg, Netherlands, New Zealand, Norway, Poland, Singapore, Spain, Sweden, Switzerland, United Kingdom.
- Recall number
- Z-2673-2020
- Recall status
- Terminated
- CDRH record ID
- 182020
- FDA record created
- 2020-07-21
- Recall event number
- 85886