Potential that the measured value may be smaller than the actual area.
Source facts
Action
Ashvins notified consignees on about 09/09/2025 via emailed letter. Consignees were instructed to cease using the freeform measurement function in the MMV until the software update has been performed, which is now available. Consignees were instructed to perform the update as soon as possible, complete the digital Field Safety Notice Acknowledgement and Receipt Form, and maintain a copy of the letter and the user manual with the device until the software update is complete.
Initiated
2025-09-09
Root cause
Software design
Distribution
US Nationwide distribution.
Recall number
Z-2673-2025
Recall status
Open, Classified
CDRH record ID
215824
FDA classified
2025-09-30
Recall event number
97509