Open original FDA record ↗Recall record
- FDA identifier
- Z-2677-2014
- Company in FDA record
- Genesys Orthopedics Systems, LLC
- FDA classification date
- Sep 22, 2014
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
The recall is being initiated because MK Precision (the Circular Lock Screw supplier) notified Genesys Spine that a portion (less than 20%) of the Circular Lock Screws in lot 11406 were manufactured using Stainless Steel in the center Post instead of the specified titanium alloy.
Source facts
- Action
- Genesys Spine sent an Urgent Medical Device Recall letter dated April 2, 2014, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to immediately examine their inventory and quarantine any affected product. Customers were instructed to notify their customers (end users) of the recall and request that they return any affected product to Genesys Spine by completing the attached Recall Tracking Form to Genesys Spine, 1250 South Capital of Texas Hwy, Building Three, Suite 600, Austin, Texas 78746. Customers with questions were instructed to call 512-381-7070.
- Initiated
- 2014-04-02
- Root cause
- Nonconforming Material/Component
- Terminated
- 2015-05-15
- Distribution
- Nationwide Distribution including NV, CA, NY, MO, TX, FL, NJ, GA, TN.
- Recall number
- Z-2677-2014
- Recall status
- Terminated
- CDRH record ID
- 126582
- FDA classified
- 2014-09-22
- Recall event number
- 67900