Open original FDA record ↗Recall record
- FDA identifier
- Z-2677-2016
- Company in FDA record
- Intuitive Surgical, Inc.
- FDA classification date
- Aug 26, 2016
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Field failures were reported related to the da Vinci Xi Surgical System EndoWrist Stapler 45 and 30 instruments having an interruption in shaft rotation and / or instrument jaws not opening.
Source facts
- Action
- Intuitive sent a Stop Use e-mail to all affected sites on July 14, 2016 to inform users of the issue prior to the commencement of a recall. E-mail described the problem and asked that all surgeons who use the stapler be informed to immediately stop use. Formal letters will follow. Medical Device Recall letters were sent on July 21, 2016. Customers were instructed to return quarantined affected product whether used or unused to Intuitive Surgical. Customers should contact customer service to initiate the standard Return Material Authorization (RMA) process by phone or email: North and South America: 800-876-1310, option 3 (6AM to 5PM PST) Email: customersupport..servicesupport@intusurg.com Europe, Middle East, Asia and Africa Phone: 800-821-2020 or +41 21 821 2020 Email: ics@intusurg.com Customers were also instructed to complete the attached Return Acknowledgement Form and submit via email to Regulatory Compliance at isi.compliance@intusurg.com or by fax to +1 (408) 523-0619. For questions regarding this recall call 408-523-2100. 10-31-16 Update: Monthly status update states that an additional 93 instruments have been identified and an additional communication will be sent to these sites in Mid November 2016,
- Initiated
- 2016-07-14
- Root cause
- Nonconforming Material/Component
- Terminated
- 2017-01-03
- Distribution
- Worldwide Distribution - US (nationwide) Internationally to Australia, Belgium, Denmark, France, Germany. Italy. Japan, Spain, Sweden, Switzerland, Taiwan, and Turkey,
- Recall number
- Z-2677-2016
- Recall status
- Terminated
- CDRH record ID
- 148514
- FDA classified
- 2016-08-26
- Recall event number
- 74811