Skip to content

Public FDA record

Z-2680-2014: The Ysio Systems with software version VC10 The Ysio enables radiographic and tomographic exposures of

Recall record Z-2680-2014 for Siemens Medical Solutions USA, Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2680-2014
Company in FDA record
Siemens Medical Solutions USA, Inc
FDA classification date
Sep 23, 2014
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Siemens discovered that an unlikely error may occur on the Ysio system with fixed detector in the wall stand. This error may result in line artifacts in the image. If these artifacts appear in the region of interest, the examination may need to be repeated.

Source facts

Action
The firm, Siemens, sent a "Safety Advisory Notice" dated August 27, 2014, to end users/customers. The notice identified the product, problem, and action to be taken. The customers were instructed to follow the instructions in the notice; promptly notify and instruct all the staff at your organization who have to be aware of this problem; forward this safety information to other organizations affected by this measure; and if the device has been sold and therefore no longer in your possession, forward to new owner. Siemens prepared a field modification of the Ysio systems with a hardware upgrade. This field modification will be available in September 2014. If you have any questions, please contact Regulatory Specialist at 610-219-4334 or Customer support at 610-219-6300.
Initiated
2014-08-27
Root cause
Under Investigation by firm
Terminated
2015-03-30
Distribution
Nationwide Distribution.
Recall number
Z-2680-2014
Recall status
Terminated
CDRH record ID
129747
FDA classified
2014-09-23
Recall event number
69148

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗