Skip to content

Public FDA record

Z-2682-2016: Philips DuraDiagnost stationary X-ray system

Recall record Z-2682-2016 for Philips Electronics North America Corporation. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2682-2016
Company in FDA record
Philips Electronics North America Corporation
FDA classification date
Sep 1, 2016
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Buttons or keys on the monitor allow the user to turn the volume setting down to 0", in which case the beep will not be heard at the conclusion of an exposure, as required 21 CFR 1020.31(j), and the Height-Adjustable Tabletop is missing the required labels [21 CFR 1010.2., 21 CFR 1010.3(a), and 21 CFR 1020.30(3)].

Source facts

Action
Philips plans to bring the products into compliance with the Federal standard at no cost to customers by implementing a field correction. 1. You will contact customers and initiate the actions to correct the failure to comply issues. 2. You will attach the warning label to maintain an audible volume, and issue a supplement to the user manual that instructs the user how to adjust the volume for the local noise environment and include a warning that the user shall maintain an audible volume. 3. You will attach the identification label to the affected Height-Adjustable Tabletop. 4. The customer notification letter which includes a statement that you will without charge, remedy the defect or bring the product into compliance. If you need any further information or support concerning this, please contact your local Philips representative or Philips Customer Service: 1-800-722-9377.
Initiated
2015-09-17
Root cause
Radiation Control for Health and Safety Act
Terminated
2017-08-31
Distribution
US Distribution to the states of : ME, NE, OR, CA, TX, NC, WA and OH.
Recall number
Z-2682-2016
Recall status
Terminated
CDRH record ID
148189
FDA classified
2016-09-01
Recall event number
74685

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗