Z-2684-2024: The VITROS XT Chemistry Products ALB-TP Slides is a single device that contains both an albumin test and

Recall record Z-2684-2024 for Ortho-Clinical Diagnostics, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2684-2024
Company in FDA record
Ortho-Clinical Diagnostics, Inc.
FDA classification date
Aug 16, 2024
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

The reason for this recall is QuidelOrtho has determined that during normal use the VITROS XT ALB-TP Slides may shed dust and debris as they travel through the MicroSlide Subsystem of the VITROS XT 3400 or VITROS XT 7600 analyzer. This dust and debris may then settle onto VITROS Na+ (Sodium) Slides as they are processed through the same MicroSlide Subsystem, potentially resulting in an increase in non-reproducible, positively, or negatively biased Na+ results. The level of bias will depend on where the dust and debris settle on the Na+ MicroSlide and may impact Na+ measurements to varying degrees inside and outside of the measuring range depending on the amount of contamination. All of which could vary from no/low bias with no patient harm, to extreme bias (falsely elevated or reduced results) potentially resulting in incorrect diagnosis; treatment; and/or a serious patient health hazard (for example, hypernatremia or hyponatremia). This issue may also potentially generate error codes preventing the analyzer from processing the MicroSlide Assays and cause delay of the patient results. All current and expired VITROS XT Chemistry Products ALB-TP slides are affected by this issue.

Source facts

Action
QuidelOrtho issued Urgent Product Correction Notification Letter to Distributors and End-Users on 10 July 2024, (Ref. CL2024-139) via FedEx overnight courier and/ or electronically using ORTHO PLUS e-Communications to customers who have been shipped VITROS XT Chemistry Products ALB-TP slides in the last 24 months. Letter states reason for recall, health risk and action to take: Until this issue is resolved, QuidelOrtho will be pausing availability of the VITROS XT ALB-TP Slides and we will shift all affected customers to the individual VITROS Chemistry Products ALB Slides and VITROS Chemistry Products TP Slides. We have verified through internal testing that the individual ALB and TP slides do not show the same behavior as the XT ALB-TP slides. Upon receipt of this notification, please perform the cleaning steps listed in Appendix 1: VITROS XT System Cleaning Focus, on your VITROS XT 3400 and/or XT 7600 System(s) to remove any dust and debris which may have accumulated in the MicroSlide Subsystem. Perform the cleaning steps listed in Appendix 1: VITROS XT System Cleaning Focus again on your VITROS XT 3400 and/or XT 7600 System(s) after your laboratory has discontinued using VITROS XT ALB-TP Slides. Customers may continue to use VITROS XT ALB-TP Slides until the replacement VITROS ALB Slides and VITROS TP Slides are received. After receipt of the replacement VITROS Slides, and on completion of any necessary crossover studies / validation, please discontinue using, render unusable, and discard your remaining inventory of VITROS XT ALB-TP Slides. cleaning of the MicroSlide Subsystem on your VITROS XT 3400 and/or VITROS XT 7600 System(s). Refer to Appendix 1: VITROS XT System Cleaning Focus for a list of specific areas which have been observed to accumulate dust and debris shed from VITROS XT ALB-TP Slides. " Prepare your laboratory s transition from using VITROS XT ALB-TP Slides to the VITROS Chemistry Products ALB Slides and VITROS TP Slides products, including
Initiated
2024-07-10
Root cause
Under Investigation by firm
Distribution
Nationwide. Foreign: Australia Belgium Bermuda Brazil Canada L3R 0B8 Chile China Colombia Denmark France Germany India Italy Japan Mexico Norway Portugal Russia Singapore 609917 Spain Sweden The Netherlands United Kingdom
Recall number
Z-2684-2024
Recall status
Open, Classified
CDRH record ID
209081
FDA classified
2024-08-16
Recall event number
94993

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗