- Recall classification
- Class II
Reason stated in the source
Interventional Radiology (IR) images are stored as JPEG2k Lossless in Merge Enterprise Archive (EA) and are not displaying correctly in RadSuite.
Source facts
- Action
- The firm, Merge, sent an "URGENT: MEDICAL DEVICE RECALL" letters dated 1/29/2016 and 5/25/2016 via e-mail or via certified mail if they did not have an e-mail address for the customers. The letters described the product, problem and actions to be taken. The customers were instructed to do the following: you must discontinue using the affected version of RadSuite or iConnect Enterprise Archive with GE AW Workstation; use the workaround listed in the letter; complete and return the enclosed form no later than June 17, 2016; ensure that all users of the product are provided with the notification; if you have further distributed this product, please identify your customers and notify them at once of this product recall Merge has a released fix available for this issue. If you have any additional questions, please send an email to recall@merge.com or call 800-724-5970 ext 8027.
- Initiated
- 2016-01-30
- Root cause
- Software design
- Terminated
- 2019-10-15
- Distribution
- US Distribution to: CO and OK.
- Recall number
- Z-2686-2016
- Recall status
- Terminated
- CDRH record ID
- 148882
- FDA record created
- 2016-08-29
- Recall event number
- 74924