Open original FDA record ↗Recall record
- FDA identifier
- Z-2687-2020
- Company in FDA record
- Sysmex America, Inc.
- FDA record creation date
- Jul 22, 2020
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Insufficient amount of antibody without an error message or alarm
Source facts
- Action
- Sysmex issued a field corrective action on July 6, 2020. The product notification states the following actions to be taken: 1. A Sysmex Service Engineer will be scheduled to adjust the probe height to correct the issue from occurring when using the Sysmex antibody vials. The PS-10 should not be used to perform any testing, other than internal validation, until the adjustment is completed. 2. Discontinue use of non-Sysmex antibody vials in the PS-10 for testing until Sysmex hascompleted the verification, unless your internal procedures ensure the required accuracy of the pipetted volume. 3. Please distribute this Product Notification as appropriate to your laboratory 4. File this Product Notification as a part of your laboratorys Quality System as required For urgent requests, call the Technical Assistance Center at 1-888-879- 7639 in the U.S., or at 1-888-679-7639 in Canada
- Initiated
- 2020-07-06
- Root cause
- Software design
- Terminated
- 2021-05-21
- Distribution
- US Nationwide distribution including in the states of AL, AZ, MD, WA, FL, IL.
- Recall number
- Z-2687-2020
- Recall status
- Terminated
- CDRH record ID
- 182272
- FDA record created
- 2020-07-22
- Recall event number
- 85959