Z-2687-2020: Sysmex PS-10 Sample Preparation System Catalog number:BQ716341 - Product Usage: is a fully automated, use

Recall record Z-2687-2020 for Sysmex America, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2687-2020
Company in FDA record
Sysmex America, Inc.
FDA record creation date
Jul 22, 2020
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Insufficient amount of antibody without an error message or alarm

Source facts

Action
Sysmex issued a field corrective action on July 6, 2020. The product notification states the following actions to be taken: 1. A Sysmex Service Engineer will be scheduled to adjust the probe height to correct the issue from occurring when using the Sysmex antibody vials. The PS-10 should not be used to perform any testing, other than internal validation, until the adjustment is completed. 2. Discontinue use of non-Sysmex antibody vials in the PS-10 for testing until Sysmex hascompleted the verification, unless your internal procedures ensure the required accuracy of the pipetted volume. 3. Please distribute this Product Notification as appropriate to your laboratory 4. File this Product Notification as a part of your laboratorys Quality System as required For urgent requests, call the Technical Assistance Center at 1-888-879- 7639 in the U.S., or at 1-888-679-7639 in Canada
Initiated
2020-07-06
Root cause
Software design
Terminated
2021-05-21
Distribution
US Nationwide distribution including in the states of AL, AZ, MD, WA, FL, IL.
Recall number
Z-2687-2020
Recall status
Terminated
CDRH record ID
182272
FDA record created
2020-07-22
Recall event number
85959

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗