Open original FDA record ↗Recall record
- FDA identifier
- Z-2687-2024
- Company in FDA record
- Preat Corp
- FDA classification date
- Aug 19, 2024
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Due to a manufacturing issue, the product has a out of specification rotation of the digital analog used to visually inspect dental abutments and restorations.
Source facts
- Action
- On 07/22/2024, the firm sent an "URGENT: MEDICAL DEVICE RECALL UPDATE" Letter via certified mail informing customers that Digital Analogs were manufactured out of specification for the rotational feature which may result in the dental restoration/implant may be slightly different than the position of the actual implant in the patient's mouth. Customers are instructed to: Product in the scope of this recall should be discarded or returned to PREAT CORPORATION. For questions contact Quality Regulatory Affairs at 224-622-7191 or email jespino@younginnovations.com
- Initiated
- 2024-06-25
- Root cause
- Process control
- Distribution
- U.S. Nationwide distribution in the states of AK, AZ, FL, GA, IL, LA, MD, MI, MN, NY, OH, OK, OR, PA, RI, TX, UT, and WA
- Recall number
- Z-2687-2024
- Recall status
- Open, Classified
- CDRH record ID
- 208842
- FDA classified
- 2024-08-19
- Recall event number
- 94933