Z-2688-2016: Invivo Sentinelle Venguard Breast MRI Auxiliary Table/Tabletop with 8/16 Channel Coil Array for GE 1.5T/3

Recall record Z-2688-2016 for Invivo Corporation. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2688-2016
Company in FDA record
Invivo Corporation
FDA record creation date
Aug 30, 2016
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

A diode was inadvertently not assembled into the system, which may result in coil overheating

Source facts

Initiated
2016-05-06
Root cause
Under Investigation by firm
Terminated
2019-08-21
Distribution
WI
Recall number
Z-2688-2016
Recall status
Terminated
CDRH record ID
146139
FDA record created
2016-08-30
Recall event number
74064

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗