Open original FDA record ↗Recall record
- FDA identifier
- Z-2688-2020
- Company in FDA record
- Siemens Medical Solutions USA, Inc
- FDA record creation date
- Jul 22, 2020
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
If the user modifies for any reason (e.g. reduction of artifacts) the original image orientation of a standard MR protocol to acquire images in a different orientation for further processing in Synthetic CT , the software does not recognize the adapted acquisition plane. This may result in images with wrong geometry. When this distortion remains unnoticed and the images are subsequently exported to a treatment planning system (TPS), an incorrect calculated radiation treatment plan cannot be excluded. The occurence of this issue is very unlikely and has never been reported so far.
Source facts
- Action
- Siemens Healthcare Diagnostics sent a letter notification to customers dated June 23, 2020. The described issue will be solved with a service pack for software versions syngo.via VB30 and syngo.via VB40. Depending on your current software version, the service pack VB30A_HF08 will be installed with update SY039/20/S or the service pack VB40B_HF01 with update SY040/20/S as soon as it is available.
- Initiated
- 2020-06-24
- Root cause
- Software design
- Terminated
- 2023-04-05
- Distribution
- US Nationwide distribution including in the states of AZ, MA, WI, FL, PA. TX, MD, LA, TX, MN, MI, NY, SC, OH, GA, CA, IL, MO, CA, TX.
- Recall number
- Z-2688-2020
- Recall status
- Terminated
- CDRH record ID
- 182403
- FDA record created
- 2020-07-22
- Recall event number
- 86000