Z-2689-2020: IntelliVue G7m Anesthesia Gas Module, Model no. 866173, Firmware Version 04.12.00, System Codes 453564477

Recall record Z-2689-2020 for Philips North America, LLC. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2689-2020
Company in FDA record
Philips North America, LLC
FDA record creation date
Jul 23, 2020
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

The device may experience an interruption of gas measurement due to a firmware issue, ceasing measurement and display of gas levels and generating one or more technical or INOP alarms.

Source facts

Action
On 6/25/2020, the firm distributed Customer Information Medical Device Correction letters to affected customers. Customers were informed that the current firmware on the 866173 G7m Gas Module needed to be updated in order to correct an issue which was causing the device to erroneously detect a premature failure of the pump. When the error occurs, the device will cease measurement and display of gas levels, and generate one or more technical or INOP alarms. Customers were instructed to ensure their staff is informed of the actions to be taken for INOP alarms, as can be found in the Instructions for Use. Local Philips Service representatives will contact customers and will update the firmware on affected devices. If you have further questions or concerns about the issue, please contact your local Philips representative or Philips Customer Support at 1-800-722-9377.
Initiated
2020-06-25
Root cause
Software design
Distribution
Worldwide distribution - US Nationwide distribution.
Recall number
Z-2689-2020
Recall status
Open, Classified
CDRH record ID
181997
FDA record created
2020-07-23
Recall event number
85795

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗