Open original FDA record ↗Recall record
- FDA identifier
- Z-2689-2024
- Company in FDA record
- Stryker Sustainability Solutions
- FDA classification date
- Aug 20, 2024
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Disposable 5 lead cable and lead wire system package that should contain lead systems, lead telemetry systems and a dual connect cable, may in fact contain, different ECG Leads, which are not approved for reprocessing, but that may have been reprocessed, which may lead to devices not performing as intended.
Source facts
- Action
- On 7/2/24, recall notices were sent to customers informing them of the following: 1) Inspect affected devices for "TruLink" branding and if found, remove them from their point of use. 2) If any of the affected products have been forwarded to additional facilities (including but not limited to distribution warehouses), please contact these facilities and communicate this recall to them. Please instruct these facilities to complete the Business Reply Form. 3) Complete and return the business reply form via email to SSSPFA@stryker.com Quality problems should be reported to the firm via phone +1-888-763-8803 or email http://www.stryker.com/productexperience
- Initiated
- 2024-07-02
- Root cause
- Labeling mix-ups
- Distribution
- US Nationwide distribution in the states of FL, WY, AZ, CA, LA, TX, MI, OR, CO, MS, TN, WI.
- Recall number
- Z-2689-2024
- Recall status
- Open, Classified
- CDRH record ID
- 209175
- FDA classified
- 2024-08-20
- Recall event number
- 95038