Skip to content

Public FDA record

Z-2690-2016: FUJIFILM Digital Mammography System Aspire HD (FDR MS-1000) with Biopsy Positioner (FDR-1000BPY) The Fuj

Recall record Z-2690-2016 for Fujifilm Medical Systems U.S.A., Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2690-2016
Company in FDA record
Fujifilm Medical Systems U.S.A., Inc.
FDA record creation date
Aug 30, 2016
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

If an exposure is aborted during a stereo biopsy workflow, an error may occur which would prevent the system from continuing to operate properly.

Source facts

Action
The firm, Fujifilm, sent a "FIELD SAFETY NOTICE" letter dated June 9, 2016 via Federal Express to its customers. The letter describes the product, problem and actions to be taken. The customers were instructed to read and follow the instructions in the letter; call your local service representative, in case an error occurs, and complete and return the Customer Feedback Form via fax (203)-251-7862. Both affected systems will be corrected via on-site visits by FMSU Customer Service Engineers. If you have any questions about this matter, please contact the Product Manager, Women's Health directly at (303)-888-4549 or call your local FUJIFILM office.
Initiated
2016-06-09
Root cause
Software design
Terminated
2016-09-19
Distribution
US Distribution to: NC and HI.
Recall number
Z-2690-2016
Recall status
Terminated
CDRH record ID
147840
FDA record created
2016-08-30
Recall event number
74525

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗