Z-2690-2017: Coulter PrepPlus 2, Part Number 378600, Product Usage: The COULTER PrepPlus 2, when used in combination

Recall record Z-2690-2017 for Beckman Coulter Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2690-2017
Company in FDA record
Beckman Coulter Inc.
FDA record creation date
Jul 7, 2017
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

When samples are prepared using the PrepPlus or PrepPlus 2 sample preparation systems and then run on a Flow cytometer, there is an imprecision in the absolute counts on only the first tube in a run.

Source facts

Action
The recalling firm initiated their recall by letter disseminated via email on 05/22/2017. The letter requested the following actions be taken: 1. Insert an empty daughter tube in place of a patient sample as the first tube in the carousel that is programmed for the addition of Flow-Count Fluorospheres. 2. Refer to the COULTER PrepPlus Operators Guide / PrepPlus 2 Instructions for Use (IFU) for instructions on adding Flow-Count Fluorospheres to daughter tubes. 3. Review all results following the flow cytometer data review instructions referenced in the associated IFU. Beckman Coulter also requested the consignee share this information with their laboratory staff and retain this notification as part of their laboratory Quality System documentation. If the consignee forwarded any of the affected product(s) listed above to another laboratory, They were asked to provide them a copy of the letter.
Initiated
2017-05-22
Root cause
Software design
Terminated
2021-03-31
Distribution
Worldwide Distribution
Recall number
Z-2690-2017
Recall status
Terminated
CDRH record ID
155842
FDA record created
2017-07-07
Recall event number
77362

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗