Z-2692-2024: FreeStyle Libre 3 Sensors, REF: 72081-01, component of the FreeStyle Libre 3 Continuous Glucose Monitorin

Recall record Z-2692-2024 for Abbott Diabetes Care, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2692-2024
Company in FDA record
Abbott Diabetes Care, Inc.
FDA classification date
Aug 27, 2024
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class I

Reason stated in the source

Users of the FreeStyle Libre 3 sensors reported situations where they were receiving erroneously high glucose results. The inaccurate higher glucose values may lead to users calculating higher insulin bolus correctional doses based on inaccurate CGM glucose data. The inaccurate higher glucose values may also contribute to missed or delayed recognition of hypoglycemia, when the sensor incorrectly reports normal or even high glucose values when the actual glucose is trending low or is low. If the hypoglycemia missed by the device occurs acutely, a user may not be able to confirm the actual glucose with a self-monitored blood glucose (BG) and/or administer prompt intervention to raise glucose levels. The over-delivery of insulin and/or the missed or delayed detection of impending or overt hypoglycemia in the worst-case scenario may lead to severe hypoglycemia/hypoglycemic crisis with significant adverse health consequences. Such consequences include central nervous system dysfunction, loss of consciousness, seizure activity and may lead to coma, permanent neurological damage, and death.

Source facts

Action
On 7/24/24, correction notices were distributed to customers, distributors, wholesalers. In addition, a news release was posted. FSL3 app users started receiving a message, with a clickable link to the notification, prior to the starting each new senor. Customers were asked to do the following: 1) To determine if your current sensor or any unused sensors you have are affected, please visit www.FreeStyleConfirm.com and select "CONFIRM SENSOR SERIAL NUMBER". 2) If you are wearing a Sensor, you can find the Serial number in the App or Reader. The Serial Number can also be found on the bottom of the Sensor Applicator label and carton. 3) If you are currently using the system and are wearing an affected sensor, please immediately discontinue use, and dispose of any affected sensors in your possession. 4) If you have symptoms that do not match the sensor glucose reading or suspect your reading may be inaccurate, check the reading by conducting a fingerstick test using any blood glucose meter. The built-in blood glucose meter in the FreeStyle Libre 3 Reader may be used to check your glucose at any time. 5) Complete and return the response form. 6) Distributors are asked to forward the communication to any customers and distribution partners they have distributed potentially impacted lots to. If you have additional questions call the firm's customer service at 1-833-815-4273, available 7 days a week from 8 a.m. to 8 p.m. Eastern Time, excluding holidays.
Initiated
2024-07-24
Root cause
Process design
Distribution
US distribution to states of: NJ, PA, OH, TX, VA, CA, NY, FL, WA, MS, MN, and WY
Recall number
Z-2692-2024
Recall status
Open, Classified
CDRH record ID
209214
FDA classified
2024-08-27
Recall event number
95047

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗