Open original FDA record ↗Recall record
- FDA identifier
- Z-2693-2011
- Company in FDA record
- Medical Components, Inc dba MedComp
- FDA classification date
- Jun 28, 2011
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 17, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
Product was packaged with the incorrect valved peelable introducer. The catheter sized is 15.5F and the introducer packaged in the kit is a 15F. The kit should contain a 16F introducer.
Source facts
- Action
- Medcomp sent a "PRODUCT ALERT" letter dated May 19, 2011 to all affected customers. The letter described the product, problem, and actions to be taken. The letter is requesting the return of all affected product. Customers are instructed to contact their customer service representative for an RGA (returned good authorization) number. If you have any questions or concerns contact Medcomp at 215-256-4201 ext. 2225.
- Initiated
- 2011-05-19
- Root cause
- Mixed-up of materials/components
- Terminated
- 2016-10-24
- Distribution
- Worldwide Distribution -- USA (nationwide) including states of AZ, CA, CO and RI and countries of Australia, Brazil, and Netherlands
- Recall number
- Z-2693-2011
- Recall status
- Terminated
- CDRH record ID
- 100757
- FDA classified
- 2011-06-28
- Recall event number
- 58975