Z-2693-2020: Accurian RF Ablation, Cannula pk, Accurian, 150mm, 20G, 10mm, CS Product Number: AC0012 GTIN / UPN: 20763

Recall record Z-2693-2020 for Medtronic Sofamor Danek USA Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2693-2020
Company in FDA record
Medtronic Sofamor Danek USA Inc
FDA record creation date
Jul 23, 2020
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Insertion difficulty of the probes. The probe was catching on a lip inside the hub of the cannula.

Source facts

Action
Medtronic initiated both a voluntary correction notice pertaining to the labeling for the affected 20G and 22Gs and removal notice pertaining to the affected 18Gs on 10 June 2020. Notice of removal for ACCURIAN RF Cannula (18G, sterile, single-use) state the following action to be taken: 1) Locate and remove the impacted product from normal storage locations. Do not use this product. 2) Complete the enclosed Customer Confirmation Form and return via email to rs.sdmriskmgt@medtronic.com. Please return this form to Medtronic even if you have no affected inventory. 3) As applicable, contact your Medtronic Sales Representative to coordinate the return and replacement of affected product. 4) Transmission of this Recall notification letter: a. This notification needs to be passed on to all those who need to be aware within your organization or to any organization where the potentially affected devices have been transferred. b. Please transfer this notice to other organizations on which this action has an impact. c. Please maintain a copy of this notice in your records. Safety Notification ACCURIAN RF Cannula (20G & 22G sterile, single-use) state the following action to be taken: Accurian RF cannula lot numbers identified may be used with adherence to the following mitigation: 1) In addition to following instructions within product IFU, the probe should be threaded straight into the center of the hub of the cannula. Do not insert a probe at an angle into the cannula. 2)If the probe catches in the cannula during insertion, remove and reinsert by threading straight into the center of the hub of the cannula.
Initiated
2020-06-08
Root cause
Under Investigation by firm
Terminated
2022-05-10
Distribution
Worldwide distribution - US Nationwide distribution and the countries of Belgium and Australia.
Recall number
Z-2693-2020
Recall status
Terminated
CDRH record ID
182120
FDA record created
2020-07-23
Recall event number
85904

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗