Z-2693-2024: IBA Proton Therapy System - PROTEUS 235- Designed to produce and deliver a proton beam for the treatment

Recall record Z-2693-2024 for Ion Beam Applications S.A.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2693-2024
Company in FDA record
Ion Beam Applications S.A.
FDA classification date
Aug 21, 2024
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Temporary deactivation of collision detection for Forte Robotic Patient Positioning System occasionally experiences issues

Source facts

Action
IBA isued Urgent Medical Device Correction letter on July 10, 2024. Letter states reason for recall, health risk and action to take: IBA would like to remind you as IBA Proton Therapy System -PROTEUS 235 user of the warnings present in your Clinical User's Guide which emphasize the importance of carefully monitoring the patients behavior during the different treatment phases. IBA would like to also remind you that the manual hand crank could be used by the system operator for the emergency release of the Patient Positioning System in the situations listed in the Clinical User's Guide and Accelerator Operator's Guide and by following the instructions described in both guides. To be signed and given back to IBA representative on site. PTS-8, PTS-10 and PTS-11 versions: Software upgrade IBA will update the software of the Positioning Management System in order that, after a collision, the request to temporarily disable the collision detection will be sent until the status of the collision detection is well disabled. This solution will be deployed at your site by March 2026. PTS-12 versions: users training and labeling change IBA will provide customer training on the system behavior in PTS-12 versions following collision detections. Additionally, IBA will update the labeling to clarify that users can also move the snout or gantry to go out collision situations. These solutions will be deployed at your site by January 2025
Initiated
2024-07-10
Root cause
Process design
Distribution
IL,LA, NJ, MI,OK, TN, TX, VA, WA Foreign: Czech Republic France Germany Italy Japan Poland Russia Sweden
Recall number
Z-2693-2024
Recall status
Open, Classified
CDRH record ID
209121
FDA classified
2024-08-21
Recall event number
95024

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗