Z-2694-2014: eye-pak 7407 Tray Support Cover REF 8065740745

Recall record Z-2694-2014 for Alcon Research, Ltd.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2694-2014
Company in FDA record
Alcon Research, Ltd.
FDA classification date
Sep 24, 2014
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

The peel pouches of the Tray Support Covers may be insufficiently sealed so that sterility cannot be assured.

Source facts

Action
Alcon Research sent an Urgent Medical Device Recall letter dated June 3, 2014, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. Please follow these important steps: 1) Review your inventory of Tray Support Covers to determine if you have any units from the affected Lots in stock. 2) Place the product aside to ensure it is not used. 3) Return the attached Response Form via fax or email to Alcon. Please return the Response Form even if you do not have any inventory from the affected lots. 4) Alcon will contact you to arrange the return of your inventory. If you have any further questions, concerns, or need assistance with obtaining replacement product, please contact Alcon Customer Service at 1-800-862-5266.
Initiated
2014-06-03
Root cause
Packaging process control
Terminated
2015-07-27
Distribution
Worldwide Distribution - Nationwide Distribution and the countries of Australia, Belgium, and Canada.
Recall number
Z-2694-2014
Recall status
Terminated
CDRH record ID
128029
FDA classified
2014-09-24
Recall event number
68487

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗