Open original FDA record ↗Recall record
- FDA identifier
- Z-2697-2020
- Company in FDA record
- Philips North America, LLC
- FDA record creation date
- Jul 23, 2020
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Some Intellivue MX700/MX800 Patient Monitors were delivered with defective power supplies. The defective power supplies might energize the metal parts of monitor if the grounding instructions were not followed, and may also cause the monitor to stop operating or prevent it from being switched on.
Source facts
- Action
- On June 30, 2020, the firm sent Urgent Medical Device Correction letters to affected customers. Customers were instructed to always ensure that the monitor is grounded during operation with mains power. Affected products should be removed from use until the power supply of the monitor has been replaced. Philips Customer Service will contact customers when a replacement power supply is ready for installation. Philips expects all replacements to be complete within six months. If you need any further information or support concerning this issue, please contact your local Philips representative or Philips Customer Support at 1-800- 722-9377.
- Initiated
- 2020-06-30
- Root cause
- Process control
- Terminated
- 2024-01-25
- Distribution
- Domestic distribution to CA, HI, MA, NY, and UT. International distribution to Costa Rica, Ireland, Lebanon, and Poland.
- Recall number
- Z-2697-2020
- Recall status
- Terminated
- CDRH record ID
- 182127
- FDA record created
- 2020-07-23
- Recall event number
- 85908