Reason stated in the source
A recent FDA inspection at the firm identified that the foil test performed during installation of the device does not have clear acceptance criteria.
Source facts
- Action
- On 6/26/2020, the firm distributed Urgent Medical Device Correction letters to affected customers. Customers were informed that a Steris service technician will contact customer facilities within 45 days to inspect their units and reperform the installation/start-up checklist. Users can continue to use their irrigator(s) prior to this inspection. Users were instructed to continue to visually inspect all instruments for visible debris prior to final high-level disinfection or sterilization, and reprocess if needed. In addition, users should continue to clean and maintain the unit before each cycle. If you have questions regarding this matter, please contact STERIS Customer Service at 1-800-548-4873, or your local STERIS Representative.
- Initiated
- 2020-06-26
- Root cause
- Process control
- Terminated
- 2020-12-22
- Distribution
- Worldwide distribution - US Nationwide distribution and the countries of Canada, Mexico, and United Arab Emirates.