Product found to be dimensionally non-conforming (slightly over dimensioned on the diameter of the Zimmer connection) and non-functional for connecting with some T-handles.
Source facts
Action
The firm received two complaints from the US field about the inta-operative malfunction with the SMR allen wrench 5mm (product code 9013.50.210) with the T-handle. Both parts involved were manufactured with the same lot #14AA311. No adverse effects for patients reported.
Initiated
2016-07-21
Root cause
Nonconforming Material/Component
Terminated
2017-01-13
Distribution
US, Italy, Australia
Recall number
Z-2707-2016
Recall status
Terminated
CDRH record ID
148532
FDA classified
2016-08-31
Recall event number
74827
Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.
Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.