Z-2707-2020: StealthStation" S7 System with Polaris Spectra Camera part number PSU 9733437 - Product Usage: are intend

Recall record Z-2707-2020 for Medtronic Navigation, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2707-2020
Company in FDA record
Medtronic Navigation, Inc.
FDA record creation date
Jul 27, 2020
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

There is a potential for specific camera units to be affected by a characterization issue during manufacturing that results in inaccurate navigation while utilizing both the active and passive trackers simultaneously. While no injuries or illnesses have been reported, If impacted cameras were to be used with both active and passive trackers simultaneously, it would result in system navigation inaccuracy of 6mm.

Source facts

Action
Urgent Medical Device Correction Letters were emailed to customers on June 5, 2020. Emails were followed by Medtronic representatives hand-delivering the recall notification letter as well as the customer confirmation form to the two affected consignees on June 12, 2020. Customers were informed that there is a potential for specific camera units to be affected by a characterization issue during manufacturing that results in inaccurate navigation while utilizing both the active and passive trackers simultaneously. Customers were informed that their local Medtronic representative would work with them to remove and replace the affected camera on your impacted system. Customers were informed that Medtronic has not identified any complaints or reports of injury related to this issue. However, if impacted cameras were to be used with both active and passive trackers simultaneously, it would result in system navigation inaccuracy of 6mm. Customers were requested to not use the affected StealthStation" S7 System unit for simultaneous active and passive navigation (such as active instrumentation, active reference frames, or a navigated microscope) until the camera is replaced on this system; to work with their Medtronic representative to replace the camera on their impacted system; and to sign and return the enclosed customer confirmation form, acknowledging receipt of this information.Customers were notified that Medtronic is communicating this information to the Food and Drug Administration (FDA). Adverse events or quality problems experienced with this product should be reported to FDA and Medtronic: Online at the FDA website (www.fda.gov) or call FDA at 1-800-FDA-1088 and via E-mail to Medtronic at rs.navtechsupport@medtronic.com or by calling 1-888-826-5603. Customers with any questions regarding the urgent medical device correction should contact Medtronic Technical Services at 1-888-826-5603.
Initiated
2020-06-05
Root cause
Nonconforming Material/Component
Terminated
2024-05-28
Distribution
US Nationwide distribution including in the states of Alabama and Massachusetts.
Recall number
Z-2707-2020
Recall status
Terminated
CDRH record ID
182108
FDA record created
2020-07-27
Recall event number
85901

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗