Open original FDA record ↗Recall record
- FDA identifier
- Z-2708-2014
- Company in FDA record
- Dako North America Inc.
- FDA classification date
- Sep 26, 2014
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Dako is recalling the Dako Link 4.0.3 software on the Dako Autostainer Link 48 because Dako has determined that an upgrade to this version of the software may cause premature rejection of reagent bottles or inadequate aspiration of reagent if the instrument has also had a preventative maintenance or waste pump replacement, and the baseplate is not completely aligned.
Source facts
- Action
- Dako North America sent an Urgent Medical Device Correction letter to all customers who were upgraded to the Dako Link 4.0.3 software that is used on the Dako Autostainer Link 48 and who also had preventative maintenance or a waste pump replacement. The letter informs the customers of the problems identified and the actions to be taken. Customers are instructed that Dako will contact customers to set up an appointment for a service visit in order to ensure that the instruments are running safely or reliably. Customers with questions are instructed to contact Dako.
- Initiated
- 2014-08-29
- Root cause
- Software change control
- Terminated
- 2015-03-03
- Distribution
- Worldwide Distribution - US Nationwide in the states of AT, AU, BE, CA, CN, DE, DK, ES, Fl, FR, GB, IE, IT, NL, NO, SE, US, DZ, GR, IS, OM, SG, UY.
- Recall number
- Z-2708-2014
- Recall status
- Terminated
- CDRH record ID
- 129827
- FDA classified
- 2014-09-26
- Recall event number
- 69191