Open original FDA record ↗Recall record
- FDA identifier
- Z-2710-2017
- Company in FDA record
- Toshiba American Medical Systems Inc
- FDA record creation date
- Jul 10, 2017
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
When Continuous Trace method is used as the tracing method on Spectral Doppler in velocity trace measurement, an incorrect value may be displayed in the ultrasound systems and in the ultrasound workstations listed below.
Source facts
- Action
- Toshiba Medical sent an Urgent Medical Device Correction letter on May 1, 2017. The consignee is instructed not to use the Continuous Trace method for cardiac velocity trace measurements to quantify the MPG values. Modified software is being provided. A Toshiba service representative will contact the consignee to make an appointment to install the modified software to fix the problem. Customers were asked to complete the form and fax it to the toll-free number shown at the top of the form. The form can also be sent by email to RAffairs@tams.com. Customers with questions were instructed to call 800-521-1968.
- Initiated
- 2017-05-01
- Root cause
- Software Manufacturing/Software Deployment
- Terminated
- 2023-10-11
- Distribution
- Nationwide Distribution
- Recall number
- Z-2710-2017
- Recall status
- Terminated
- CDRH record ID
- 156748
- FDA record created
- 2017-07-10
- Recall event number
- 77599