Skip to content

Public FDA record

Z-2712-2017: APLIO i-series: a) APLIO i700 TUS AI700 b) APLIO i800 TUS AI800 diagnostic ultrasound system

Recall record Z-2712-2017 for Toshiba American Medical Systems Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2712-2017
Company in FDA record
Toshiba American Medical Systems Inc
FDA record creation date
Jul 10, 2017
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

When Continuous Trace method is used as the tracing method on Spectral Doppler in velocity trace measurement, an incorrect value may be displayed in the ultrasound systems and in the ultrasound workstations listed below.

Source facts

Action
Toshiba Medical sent an Urgent Medical Device Correction letter on May 1, 2017. The consignee is instructed not to use the Continuous Trace method for cardiac velocity trace measurements to quantify the MPG values. Modified software is being provided. A Toshiba service representative will contact the consignee to make an appointment to install the modified software to fix the problem. Customers were asked to complete the form and fax it to the toll-free number shown at the top of the form. The form can also be sent by email to RAffairs@tams.com. Customers with questions were instructed to call 800-521-1968.
Initiated
2017-05-01
Root cause
Software Manufacturing/Software Deployment
Terminated
2023-10-11
Distribution
Nationwide Distribution
Recall number
Z-2712-2017
Recall status
Terminated
CDRH record ID
156751
FDA record created
2017-07-10
Recall event number
77599

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗