Open original FDA record ↗Recall record
- FDA identifier
- Z-2713-2016
- Company in FDA record
- Cardinal Health
- FDA record creation date
- Sep 2, 2016
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Arterial Blood Gas (ABG) Kit- Missing Safety Component. The ABG Kit(s) have been identified as missing the locking clasp component. CareFusion/BD has identified a potential customer/user safety risk associated with the missing locking clasp component.
Source facts
- Action
- CareFusion/BD sent out a FIELD SAFETY NOTICE of the field correction in writing on June 27, 2016. The letter described the nature of the recall and provided actions to be performed. Carefusion/BD directed customers to take the following actions: Step #1: Inspect current inventory on-hand to identify the presence of the affected Arterial Blood Gas Sampling Kit(s). Step #2: Complete the Customer Response Form by checking all appropriate boxes and providing all accurate information as required. Step #3: Destroy all affected product(s) in-stock in accordance with their facility's destruction protocol. Step #4: Return the Customer Response Form to GMB-GLB-ALFieldActions@CareFusion.com Step #5: If the customers want to obtain credit and/or replacement of destroyed product(s) they are asked to call their CareFusion/BD Customer Support Department at 800.323.9088 (Option #1 ). For any additional questions and support concerning the voluntary Field Safety Corrective Action, customers are asked to call (847)362.9303.
- Initiated
- 2016-06-27
- Root cause
- Under Investigation by firm
- Terminated
- 2018-03-02
- Distribution
- Worldwide Distribution - US (nationwide) and Canada.
- Recall number
- Z-2713-2016
- Recall status
- Terminated
- CDRH record ID
- 148260
- FDA record created
- 2016-09-02
- Recall event number
- 74706