Z-2713-2026: GSP Neonatal GALT Kit. Product Codes: 3303-001U, 3303-0010. in vitro diagnostic. Used in FLUORESCENT P

Recall record Z-2713-2026 for WALLAC OY. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2713-2026
Company in FDA record
WALLAC OY
FDA classification date
Jul 9, 2026
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Reduction in the on-board stability of the GSP GALT assay. This may increase the likelihood of run-QC failures and consequently delay the reporting of patient results.

Source facts

Action
Wallac Oy, a Revvity company, notified consignees on about 05/18/2026 via emailed "URGENT: MEDCIAL DEVICE RECALL" letter. Consignees were instructed to inspect inventory for the affected lots and dispose of affected lots according to local requirements. If consignees do not have another GSP Neonatal GALT kit lot at hand, the newborn screening for classical galactosemia may be continued, with following instructions: - Reconstitute GALT Substrate Reagent by adding 2.8 mL of GALT Assay Buffer and invert gently for 20-30 minutes (note: slightly extended dissolving time). - Reagent cassettes should preferably be used within 12 hours (note: reduced on-board storage time). - Load reagent cassette to GSP without additional delay after reconstitution of Substrate Reagent (one reagent cassette is sufficient for measurement of four full plates (384 wells). - If possible, prioritize GALT plates first if multiple analytes/kits are run within the same instrument. Avoid exceeding the on-board storage time of 12 hours for the reagent cassette. - Run calibration curve on each plate and evaluate the results against the plate-specific curve (note: do not use stored calibration curve). These precautions are aimed at minimizing the fluctuation of the GALT activity results between assays/plates. Consignees were also requested to complete and return the response form provided. Revvity will replace the quantity of excessive kit usage and the affected GSP Neonatal GALT kits disposed from inventory upon receipt of the response form.
Initiated
2026-05-18
Root cause
Nonconforming Material/Component
Distribution
Worldwide - US Nationwide distribution in the states and territories of AL, AR, AZ, CO, FL, GA, IL, IN, KY, LA, MA, MD, MN, MO, NJ, OR, PR, SC, TN, TX, VA, WV and the countries of Austria, Belgium, Egypt, France, Germany, Hong Kong, Iraq, Italy, Netherlands, Oman, Philippines, Saudi Arabia, Singapore, Spain, Thailand, United Arab Emirates, Vietnam.
Recall number
Z-2713-2026
Recall status
Open, Classified
CDRH record ID
220725
FDA classified
2026-07-09
Recall event number
99065

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗